J-STAR
OMP 5 CALIBRATION INSTRUCTIONS
To
calibrate the OMP 5 scale indicator you must start by entering the setup mode.
This is done by pressing the [NET/GROSS] key in conjunction with the [ON] key
for three seconds or until the small cal enunciator shows up at the top of
the display. The first thing that will be displayed will be tc as shown
below.
To change
displayed values simply press the [NET/GROSS] key until the correct value is
displayed. To enter the value and proceed to the next step press the [ON] key.
|
1 |
tc |
time
constant |
4 |
|
2 |
out |
output
units |
LB |
|
3 |
cnts |
display
count |
10 |
|
4 |
stabl |
stability |
off |
|
5 |
otrac |
zero
tracking |
off |
|
6 |
tr4 |
transmitter
setup |
net |
|
7 |
AutoP |
spare
option |
off |
|
8 |
in |
input units |
LB |
|
9 |
orng |
over range
value |
42000 |
|
10 |
bal |
balance |
|
Balance the
scale using the [NET/GROSS] key and the [TARE] key. At this time add a known
weight to the scale. (You must add at least 10% of the gross capacity or
4000LBS.)
Use the
[NET/GROSS] key to change the value of the flashing digit. And use the [TARE]
key to change the digit that is flashing.
When the
correct number has been entered press the [ON] key to enter the normal weighing
mode.