CHANGING CALIBRATION BETWEEN LOADCELLS AND WEIGHBARS FOR A OMNI 510 SCALE
1.With the scale turned on press the [OMSI] key.
2. (SEC id) will be displayed.
3. Using the [NET/GROSS] key and the [TARE] key enter in the number 1 digit six times.
The [TARE] key is the increment advance, and the [NET/GROSS] key is the position
Advance. With the display showing (SEC id) press the [TARE] key. Display now shows
‘0’ press the [TARE] key to change it to a number ‘1’. Then press [NET/GROSS] key, display
now shows’10’. Press [TARE] key to change that zero to a ‘1’. Display now shows ‘11’.
Press [NET/GROSS] key, display shows ‘110’. Press [TARE] key to change zero to a ‘1’.
Keep repeating until all six number 1’s are displayed.
4. With all six number 1’s displayed press the [PRINT] key one time. The print key has now
became the ENTER key.
5. (Uid) will be displayed.
6. Using the [OMSI] key press it 9 times until (AutoC) is displayed.
7. At (AutoC) press and hold the [PRINT] key ,then press the [OMSI] key, unit will beep and then
change to a number, let go of both keys. Number displayed will be a ‘0’ or a ‘9’. At this time
if changing calibration from loadcells to weighbars press the [TARE] key until the number displayed
is a ‘2’. If changing calibration from weighbars to loadcells press [TARE] key until number
displayed is ‘1’. With the desired calibration number now showing on the display, press the
[PRINT] key. Scale will now display(Uid). Press [NET/GROSS] key to return scale to weighing
mode.
Calibration mode.
AutoC 1: Preset calibration for 4 30k loadcells
with a system capacity of 40,000 lbs.
AutoC 2: Preset calibration for 4 10k weighbars with a system capacity of 40,000 lbs.