CALIBRATION FOR EZ
2000V OR EZ 2000 USING TEST WEIGHTS
- Power up the
scale and allow to warm up for 5 minutes.
- Press and hold
the [NET/GROSS] key, then press the [ON] key until the scrolling message
appears: PRESS SELECT FOR- MENU 1 –MENU- 2 MENU- 3- CALIBRATION- EXIT
THEN PRESS ON
- Press the
[SELECT] key until CALIB appears on display, then press the [ON] key.
- T CALB will appear, and
then change to ON.
- Press the [ON]
key. ZERO will appear. Press the [NET/GROSS] key then press the
[ZERO] key to zero balance the scale.
- Press the [ON]
key, ADD WEIGHT will appear. LOWER WEIGHT RACK ONTO MIXER.
- Gross weight of
the test rack will be displayed. Press the [ON] key.
- CAL will be
displayed, then the weight of test rack will be
displayed with the first digit flashing.
- To change the
calibration number the [NET/GROSS] key will act as the VALUE INCREMENT key
and the [TARE] key will change the position of the flashing number.
- When the desired
number is entered into the scale press the [ON] key
- GOOD will be
displayed, then the scrolling message in step #2
will be displayed.
- Press the
[SELECT] key until EXIT appears, then
press the [ON] key. Scale will now display new calibrated gross weight. Remove
test weight rack. Scale should display zero. Turn scale off.