EATON 625 CALIBRATION
INSTRUCTIONS
1. Turn the scale on by
pressing the "ON" button and allow 5 minutes for warm up.
2. Zero balance the scale
by using the coarse balance control located inside the scale indicator. Adjust
the scale as close to zero as possible. (Figure 1)
3. Next press and hold
the "TOTAL LOAD" and "ZERO" buttons for three seconds to
zero the remaining value displayed on the indicator.
4. Add a known weight to
the scale.
5. Adjust the calibration
pot (Figure 2) until the known weight is displayed on the indicator.
6. Remove the weight from
the scale.
7. If the display does
not read zero after removing the weight, repeat steps 2 through 6 until the
scale settles at zero when the weight is removed.