OMNI 504 SCALE INDICATOR
CHANGE WEIGHING METHOD
LOAD CELLS -
WEIGHBARS
1.
With
the scale turned on, press the [OMSI] key.
2.
(SEC id) will be displayed.
3.
Using
the keypad, press the 1 six times (- - - - - -) will be displayed.
4.
Press
the [ENTER] key. (Uid) will be displayed.
5.
Press
the [OMSI] key 9 times until (AutoC) is displayed.
6.
At
(AutoC), press and hold the [PRINT] key, then press
and hold the [OMSI] key. The unit will beep, then
change to a number. Release of both keys. The number displayed will be a 0 or a
9. If changing from load cells to weighbars, press 2. press
the [ENTER] key two times. (0-CAL) will now be displayed. Press the [ZERO] key.
(Uid) will be displayed. Press the [NET/GROSS] key to
return to weighing mode. The scale is now calibrated for weighbars. If changing
from weighbars to load cells, enter a 1 at (AutoC)
and follow the same procedure just described.
|
AutoC 0: |
Zero is used by the AutoC
selection in the OMSI mode to enter the manual calibration mode. |
|
AutoC 1: |
Preset calibration for four 30K load cells
with a system capacity of 40,000 lbs. |